Dimensions and weight
Height 44 cm / 17.3 in
Width 91 cm / 35.8 in
Depth 72 cm / 28.3 in
Weight 60 kg / 132 lb
Temperature and humidity
System operating temperature: from +19°C (+66.2°F) to +26°C (+78.8°F)
Humidity conditions: relative humidity of 50%
Maximum for temperatures up to +26°C (+78.8°F)
Maximum relative humidity 50%
Sample volume
Automatic cycle 75 µL
Power supply and electrical consumption
Power supply 100 V to 240 V (+- 10%), 50 Hz to 60 Hz, maximum consumtion 200 VA
Reagents
ABX Diluent
May Grunwald
Giemsa
Buffer (pH 6.8 or pH 7.2)
Ethanol
Maintenance
Automatic cycles
Consumable
Slide frosted rounded corner, beveled edges.
Dimensions: 75 mm (+1/-0) x 25 mm (+1 / -0) x 1 mm (+/-0.5)
Smearing tape: 1 Strip for 6000 smears
Spreading profiles
Defined and adjustable by
• Volume of drop deposited
• Degree of tilt
• Smearing speed
Standard slide-staining protocol
May Grunwald Giemsa and Wright
Open slide-staining protocol
Wright Giemsa
Capacity
160 slide loader (continuous loading)
Loader for 8 clide-holding racks (continuous loading)
20 slides per rack
Programming reflex tests (Yumizen P8000)
Results (parameters, alarms, delta check, etc.)
Demographics (service, prescriber, age, etc.)
Activation mode
Manual (worklist)
Automatic (reflex test)
Identification
Automatic three-line printing on the slide with indelible ink
OCR (optical character recognitaion) or 2D barcodes
Staining in manual mode
The slide-staining module can be used in isolated manual mode for staining external slides.